Settings Menu - Riester ri-vital Instrucciones De Uso

Ocultar thumbs Ver también para ri-vital:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6.2 Settings menu

After pressing the On/Off button the main screen appears after 5 seconds, you arrive in
the Settings menu by pressing the Setup button.
Pulse signal
Activate NIBP mode
Cuff pressure
Exit
The „NEO" setting must only be used for neonates and infants with an arm
circumference of less than 13 cm.
Risks with incorrect settings for infants measurement:
If a blood pressure measurement is carried out on a newborn infant with the wrong
setting (adult), the ri-vital recognizes the wrong setting by the connected neonatal cuff
and immediately stops the measurement or inflating. There are no risks arising from
incorrect adjustment.
The permanent presetting is only used for easier operation in certain surroundings, e.g.
the inflation pressure for patients who are mostly hypertonic can be set higher from the
outset.
The settings can be highlighted yellow with the
Setup button.
To exit the menu, press the
confirm the settings with the Setup button.
ri-vital
OPERATING CONTROLS / BASIC SETTINGS
56
(adult oder neonate)
menu button until the End-Back setting is reached and
I
menu buttons and changed with the

Publicidad

Tabla de contenido
loading

Tabla de contenido