Current surge
Voltage drops
Test
Voltage drops
Voltage drops
Voltage drops
Power frequency 3A/m
ri-vital
MANUFACTURER'S DATA REGARDING EMC
82
Output
Output
+/- 1kV difference
+/-1kV difference
+/- 2kV common
+/- 2kV common
mode IEC 61000-4-5
mode
<5% Un(> 95%
drop in Ut)
see test step
Test step
Classification
40% Un (60%
see test step
drpo in Ut) with
five full changes.
70% Un (30%
see test step
drop in Ut) with
25 full changes
<5% Un (60%
see test step
drop in Ut) in
five seconds
3A/m
IEC 61000-4-4
The mains power should correspond to the power in
a typical hospital or business environment.
The mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then an
uninterrupted power supply must be used.
Comment
The mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then an
uninterrupted power supply must be used.
The mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then an
uninterrupted power supply must be used.
The mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then
an uninterrupted power supply must be used. IEC
61000-4-11
In the event of operating problems, the device must
be placed at a greater distance to the source of the
magnetic fields. IEC 61000-4-8
NOTE: Un is the mains voltage for the test device