Spare parts/accessories (not included in the scope of delivery)
Fig.
Item
4
–
–
–
5 Preparation for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
INFORMATION
► Not all of the materials may be available in your country. In this case,
please contact the local branch of the manufacturer to obtain informa
tion on alternative materials.
5.1 Installing the foot adapter
► Install the foot adapter on the prosthetic foot according to its instructions
for use.
5.2 Assembling the ankle block (exoskeletal design)
Required materials : Screw connection (see Page 14), 710D4 torque
>
wrench, 40 grit sandpaper, 617H21 sealing resin, 617P37 hardener
powder, suitable solvent (e.g. 634A28), only for 1WR95: 636W17 plastic
adhesive with 636W26 cross-linking agent
1) Insert the threaded socket of the screw connection in the proximal side
of the hole in the ankle block.
2) Insert the screw on the foot sole side in the prosthetic foot and screw the
ankle block to the prosthetic foot (mounting torque: 30 Nm).
3) Conduct the fitting of the ankle block.
4) Loosen the screw connection between the prosthetic foot and the ankle
block.
5) Sand the ankle block according to the contour of the prosthetic foot.
6) Roughen the connection surface of the prosthetic foot in an a – p direc
tion using the sandpaper.
7) Remove the dust from the sanded surfaces.
Designation
Screw connection for ankle block
(exoskeletal design)
Ottobock special adhesive with
hardener
Reference number
2Z22=M10
636W28 or 636W18 and
636W19
15