PXP Series Balance
8. Once the span calibration is completed successfully,
"Calibration done" will display.
Press any button to return to the previous screen.
5.2.5 Linearity Calibration
Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
For detailed calibration mass information, please see the specification tables in section 9.1.
With no load on the scale, press Linearity Calibration to begin the process.
The balance captures the zero point, and then prompts for the next weight.
Continue to follow the instructions on the display until the calibration is completed.
Steps for linearity calibration
1. Press and hold the Tare / Menu-Cal button, and the
Calibration Menu will display.
2. Press Yes to enter the Calibration Menu. To change
the calibration mode, press No until Linearity Cal
(linearity calibration) is displayed.
3. Press Yes to begin the linearity calibration.
EN-17