EN-18
4. Calibration masses value will be shown in the screen.
After the display shows "Place weights" and "50.0000
g", place weight(s) of 50 g on the pan for calibration.
5. Remove the weight(s) of 50 g from the pan. After a
while, "100.0000 g" will display. Place weight(s) of 100
g on the pan for calibration.
6. Once the linearity calibration is completed successfully,
"Linearity done" will display.
Press any button to return to the previous screen.
5.3 Balance Setup
Enter this sub-menu to customize balance functionality.
Note: The factory default settings are shown below in bold.
5.3.1 Password
Set or clear the password.
Off = disabled
On = enabled
When the Password is set On and the password is not correctly entered, the menu settings are affected as
follows:
Calibration Menu:
Auto Calibration, InCal Adjustment, Span Calibration and Linearity Calibration are locked at the current settings.
Setup Menu:
Language, Filter level, Auto Zero Tracking, Auto Tare, Graduations, Date Format, Date Setting, Time Format and
Time Setting are locked at the current settings.
Units Menu:
All settings are locked.
Print Menu:
All settings are locked.
GLP Menu:
All settings are locked.
PXP Series Balance