PXP Series Balance
MAIN APPLICATION SCREEN
Net (NET)
Stability (*)
Negative (-)
Centre of Zero (>0<)
Pre-tare (PT)
Instructional Messages
3.2
Principal Functions and Main Menu
Weighing:
First press Zero to set the display to zero. Place an object on the pan. The display indicates
the gross weight.
Taring:
With no load on the pan, press
the pan and press Tare. Add material to the container and its net weight is displayed. After
the container and the objects are removed, the load will be displayed as a negative number.
Press Tare to clear.
Zero:
Press
Dot-matrix Display:
The relevant data in the specific application mode are shown in the dot-matrix display area.
3.3
Overview of Parts and Features – Draft Shield Models
Adjustable
Foot
3.4
Overview of Parts and Features – Non-Draft Shield Models
Adjustable Foot
Zero
to zero the balance.
Pan
Level Indicator
Adjustable
Foot
Pan
Adjustable Foot
Zero
to set the display to zero. Place an empty container on
USB
Device
Security Switch RS232
USB
Security Switch
Device
EN-7
Result Field:
Information varies
by application
Unit
Power Input
Power Input
RS232