10. Technical data of the measuring accessories
4 mm Safety test leads ATL 2
-
Standard: EN 61010-031,
-
Maximum rated voltage to earth ( ) and measuring category:
1000 V CAT III, 600 V CAT IV,
-
Maximum rated current: 10 A,
-
Protective class II ( ), continuous double or reinforced insulation,
-
Contamination class: 2,
-
Length: 1.4 m, AWG 18,
-
Environmental conditions:
Maximum barometric elevation for making measurements: 2000 m,
Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %
-
Only use the test leads if in perfect condition and according to this manual,
since the protection provided could otherwise be impaired.
-
Throw the test leads out if the insulation is damaged or if there is a break in
the cable/ plug.
-
Do not touch the bare contact tips of the test leads. Only grab the area
appropriate for hands!
-
Insert the angled terminals in the testing or measuring device.
11. Environmental notice
At the end of the product's useful life, please dispose of it at
appropriate collection points provided in your country.
11/ 2007
BENNING CM 2
16