Make sure to observe any operating conditions and safety instructions stipulated by the man
ufacturer of the implant.
CAUTION
Distance to HF communication devices is too small (e.g. mobile phones, Bluetooth
devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
►
Therefore, keeping the following minimum distances to these HF communication devices is
recommended:
• Mobile phone GSM 850/GSM 900: 0.99 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points,...): 0.22 m
• Bluetooth devices (third-party products not approved by Ottobock): 0.22 m
CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention
systems, metal detectors)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
►
Avoid remaining in the vicinity of visible or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters,
transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
►
When walking through theft prevention systems, body scanners or metal detectors, watch
for unexpected behaviour of the product.
CAUTION
Remaining in areas outside the allowable temperature range
Injury due to faulty control or malfunction of the product.
►
Avoid remaining in areas outside the allowable temperature range (see Page 13).
CAUTION
Penetration of product with dirt and humidity
Injury due to unexpected product behaviour or malfunction.
►
Ensure that neither solid particles nor liquids can penetrate into the product.
CAUTION
Independent manipulation of the product
Injury due to product damage or malfunction.
►
Manipulations to the product other than the tasks described in these instructions for use are
not permitted.
►
The product and any damaged components may only be opened and repaired by authorised,
qualified Ottobock personnel.
5 Scope of Delivery
•
1 pc 13E100 AnalogAdapter
Ottobock | 11