1.1. General

The Sonopuls 190 is a unit for continuous and pulsed
ultrasound therapy at a set frequency of 1 MHz.
All functions of the unit are controlled and moni-
tored by a built-in microprocessor, ensuring a high
degree of reliability and safety.

1.2. Therapeutic possibilities

By using continuous and pulsed ultrasound it is
possible to achieve muscle relaxation and pain re-
duction and also to improve the blood perfusion
and recovery ability of tissues.
The Sonopuls 190 cannot be connected to an elec-
trotherapy unit. It is therefore not possible to apply
combinationtherapy with the Sonopuls190.

1.3. Treatment head

The large treatment head is permanent connected
to the unit. The head has excellent beam character-
istics, fully meeting the requirements of the existing
standards. The treatment head is suitable for
subaqual treatments. The connection of a second,
small treatment head, is not possible.

1.4. Efficiency and safety

During treatment the Sonopuls 190 constantly
checks the contact control. As soon as the acous-
tic contact is poor the therapist is warned by means
of a audio signal (if turned on, see chapter "opera-
tion"), the intensity will be reduced automatically
and the treatment timer will stop.
The contact control in combination with the excel-
lent beam characteristics, maintains the quality of
the treatment. Maximum safety is inherent in the
design of the unit. Both the electrical safety and
the ultrasound safety of the Sonopuls 190 comply
fully with the IEC safety standards.

1.5. Programs

In the Sonopuls 190 9 easy adjustable treatment
suggestions are available which serve as a guide-
line for treatments. A free programmable start up
program ensures that, after switching on, the unit
automatically sets itself to the most frequently used
therapy type. For setting of the start up program
and an overview of the pre-programmed programs
we refer to chapter 9 "Operation" of this manual.

1.6. Finally

You have made a wise choice in selecting the
Sonopuls 190. We are confident that your unit will
continue to give satisfaction over many years of
use. Nevertheless, if you have any queries or sug-
gestions, please contact your authorised distributor.

1. INTRODUCTION

2. PRELIMINARY NOTES

It is important that you read these operating in-
structions carefully before using the Sonopuls 190.
Please make sure that these instructions are avail-
able to all personnel who operate the equipment.
For the various applications of this apparatus,
please refer to our therapy book "Ultrasound
Therapy" reference number 1482.762.
Pay attention to the following before using the
Sonopuls 190:
1. Keep yourself informed of the contra-indications
(see page 19).
2. The apparatus may not be used in close prox-
imity (i.e. less than 2 metres) to short-wave or
microwave equipment.
3. The apparatus may not be used in so-called
"wet-rooms" (hydrotherapy rooms).
The manufacturer cannot be held responsible for the
results of using this apparatus for any purposes other
than described in these operating instructions.

3. INSTALLATION AND SAFETY

3.1. Installation

• Do not install the unit in a location near to a heat
source such as a radiator.
• Avoid exposure to direct sunlight, rain, excessive
dust, dampness, mechanical vibrations and
shocks.
• Cordless telephones should not be used in close
proximity of the unit.
• This unit should not be used in so-called "wet-
rooms" (hydrotherapy rooms). The unit has to be
installed in such way that liquid cannot enter.
• Always use the by Enraf-Nonius supplied mains
adapter (type ENB-1530).
Should any liquid ingress into the housing of this equip-
ment, unplug the unit from the wall socket (if connected)
and have it checked by an authorised expert.

3.2. Connection

• Mains supply connections must comply with the
national requirements regarding medical rooms.
• Before connection of this apparatus to the sup-
ply, check that the mains voltage and frequency
stated on the type plate of the mains adapter
corresponds to that of the mains supply.
• The mains adapter is a part of the supply circuit
on which the safety of the equipment is based.
As a consequence, the certificates (CB and
GOST) of the Sonopuls 190 are only valid in
combination with this particular type of mains
adapter (ENB-1530).