Preventative Maintenance - Stryker 1037 Manual De Uso

Camilla para traumatismos
Ocultar thumbs Ver también para 1037:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 107
At a minimum, preventative maintenance should be performed annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
All fasteners secure
Siderails move and latch properly
Engage brake pedal and push on the stretcher to ensure all casters lock securely
Steer function working properly
All casters secure and swivel properly
Inspect each caster and remove any wax or debris which may have collected on the caster or braking
mechanism
Body restraints working properly
I.V. pole intact and operating properly
Oxygen bottle holder intact and operating properly
Fowler operating and latching properly
Trendelenburg/reverse Trendelenburg operating properly
No rips or cracks in mattress cover
Transfer boards intact and operating properly
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks holding properly
Hydraulic drop rate set properly
Hydraulic oil level sufficient
Lubricate where required
X-ray tray slide tracks cleaned of dust and debris
Accessories and mounting hardware in good condition and working properly
Product Serial Number:
Completed by: ____________________________________________ Date: _____________________
www.stryker.com

Preventative Maintenance

1037-009-005 REV B
Return To Table of Contents
1-33
English

Publicidad

Tabla de contenido
loading

Tabla de contenido