4.2
Connecting the electrodes
•
Use always demineralized water with vacuum electrodes to
avoid lime deposits in the watertank, tubes and sponges.
Add a saline solution to improve the electrical conduction.
•
Use moist sponges only. Too dry sponges can cause a bad
electrical contact and burn the skin.
•
Do not use vacuum electrodes with DC current. The DC
current causes damage to the vacuum cups by ionization.
4.2.1
Using four vacuum electrodes
1. If present disconnect the vaginal,
anal and rectal stimulation
probes.
2. Connect the vacuum electrodes
to the vacuum hoses.
3. Connect the four vacuum hoses.
Select two cables with the same
hose colour for each channel.
1
Connect the red connectors
from the vacuum hoses to
the output connectors with
the red dot.
2
Connect the black connectors from the vacuum hoses to the
output connectors with the black dot.
4. Moisten the round sponges.
5. Put the sponges in the vacuum electrodes.
6. Turn the vacuum intensity regulator to
suction force.
7. Place the vacuum electrodes on the part of the body that must be
treated. The vacuum electrodes stay in place by the underpressure.
Too high suction cause patient discomfort.
8. Turn the pulse vacuum suction power regulator to
comfort.
9. Select the stimulaton signal on the vacuum output
button. The LED vacuum channel B is lit.
Vaco 200
and regulate the desired
14
for patient
with the push
B