Guidance and manufacturer's declaration - electromagnetic immunity
The 200-series devices are intended for use in the electromagnetic
environment specified below. The customer or the user of a 200-series device
should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Electrostatic
±6 kV contact
Discharge
±8 kV air
(ESD)
IEC 61000-4-
2
Electrical fast
±2 kV for
transient/burst
power supply
lines
IEC 61000-4-
±1 kV for
4
input/output
lines
Surge
±1 kV
differential
IEC 61000-4-
mode
5
±2 kV
common mode
Voltage dips,
<5% U
short
(>95% dip in
interruptions
U
) for
T
and voltage
0,5 cycle
variations on
40% U
power supply
dip in U
input lines
5 cycles
IEC 61000-4-
11
70% U
dip in U
25 cycles
<5% U
(>95% dip in
U
) for 5 sec
T
Power
3 A/m
frequency (50/
60 Hz)
magnetic field
IEC 61000-4-
8
NOTE
U
is the a.c. mains voltage prior to application of the test level
T
Vaco 200
Compliance level Electromagnetic environment -
±6 kV
contact /
±8 kV air
No loss of
performance
±2 kV power / ±1
kV I/O
No loss of
performance
±1 kV diff. /
±2 kV comm.
No loss of
performance
U
- 100% (0,5
T
T
period)
No loss of
performance
(60%
U
- 60% (5
T
T
) for
periods)
T
No loss of
performance
(30%
U
- 30%
T
T
) for
(25 periods)
T
No loss of
performance
U
- 100%
T
T
(5 seconds)
Device resets to a
safe state. (60601-
1
§ 49.2)
Not applicable
190
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity must be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of a 200-series device requires
continued operation during power
mains interruptions, it is
recommended that the 200-series
device be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.