4.2.2
Using two vacuum electrodes
1. If present disconnect the vaginal,
anal and rectal stimulation
probes.
2. Connect two vacuum electrodes
to the vacuum hoses. Select two
cables with the same hose
colour.
3. Connect the red connector from
the vacuum hose to the channel
A output connector with the red
dot.
4. Connect the black connector from the vacuum hose to the channel
A output connector with the black dot.
5. Connect a vacuum hose for a closed vacuum system to make a
short-circuit across the channel B output connectors.
6. Moisten the round sponges.
7. Put the sponges in the vacuum electrodes.
8. Turn the vacuum intensity regulator to
suction force.
9. Place the vacuum electrodes on the part of the body that must be
treated. The vacuum electrodes stay in place by the underpressure.
Too high suction cause patient discomfort.
10. Turn the pulse vacuum suction power regulator to
patient comfort.
4.2.3
Using traditional electrodes
1. Prepare the traditional
electrodes. See the User
Manual of the ET device.
2. Connect the two-ply cable to
the corresponding output
connector
Vaco 200.
Vaco 200
or
of the
A
B
and regulate the desired
15
for the