1.2.2

Electrical safety

Only use the equipment in an area with facilities that meet
the applicable legal regulations.
Connect the equipment to an outlet with a protective earth
terminal. The outlet must meet the locally applicable
requirements for medical areas.
1.2.3
Prevention of explosion
Do not use the equipment in an area where combustible
gases or vapours are present.
Switch off the equipment when it is not used.
1.2.4
Electro Magnetic Compatibility
Medical electrical equipment requires special precautions
for Electro Magnetic Compatibility (EMC). Follow the
instructions for the installation of the equipment. See §2.
Do not use mobile telephones or other radio, shortwave, or
microwave equipment in the vicinity of the equipment. This
kind of equipment can cause disturbances.
When the device operates in the vicinity of equipment
other than those mentioned in §2.3, verify that the Vaco
200 operates normally.
Only use the accompanying accessories that are supplied
by GymnaUniphy. See §7.4 and §7.5.
Other accessories can lead to an increased emission or a
reduced immunity.
1.2.5
Accessories
Check the electrode cables and the electrodes at least once
a month. Check whether the insulation is still intact. See
§5.1.
Use always demineralized water with vacuum electrodes to
avoid lime deposits in the watertank, the tubes and the
sponges.
Use only moisten sponges. Too dry sponges can cause
burning of the skin under the electrodes.
Do not use vacuum electrodes with DC currents. The DC
currents cause damage to the vacuum cups by ionization.
Vaco 200
8