10.4.3 Recommended separation distances between portable and mobile RF
communications equipment and the Sonopuls 190
The Sonopuls 190 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Sonopuls 190 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Sonopuls 190 as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
Separation distance according to frequency of transmitter
output power
[ m ]
of transmitter
150 kHz to 80 MHz
d = [3,5/V
[ W ]
0,01
0,12
0,1
0,37
1
1,17
10
3,69
100
11,67
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people
80 MHz to 800
]√P
MHz
1
d = [3,5/E
0,12
0,37
1,17
3,69
11,67
800 MHz to 2,5 GHz
d = [7/E
]√P
1
]√P
1
0,23
0,74
2,33
7,38
23,33
53