The following tables outline the test levels of the tests to be carried out. The higher value applies if
there is a difference between the operating environments listed in the tables below.
Table 1 – Electromagnetic emissions for all devices and systems
Interference measurements
HF emissions according to
CISPR 11
HF emissions according to
CISPR 11
Harmonics according to
IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
Table 4 - casing
Phenomenon
Electrostatic dis
charge
High frequency elec
tromagnetic fields
High frequency elec
tromagnetic fields in
the immediate vicinity
of wireless communic
ation devices
Magnetic fields with
rated power frequen
cies
Table 5 – alternating current gate for the supply input
Phenomenon
Electrical fast transi
ents/bursts
Surges
Line against line
Compliance
Group 1
Class B
N/A
N/A
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
EMC basic standard
or
test procedure
IEC 61000-4-4
IEC 61000-4-5
Electromagnetic environ
ment directive
The product uses HF energy
exclusively for its internal func
tioning. Its HF emissions are
therefore very low, and inter
ference with neighbouring
electronic devices is unlikely.
The product is intended for
use in other than residential
facilities and in facilities con
nected directly to a public low-
voltage network that also sup
plies residential buildings.
Interference immunity test level
Professional health
care facility
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
See Table 9
30 A/m
50 Hz or 60 Hz
Interference immunity test level
Professional health
care facility
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
Environments in
areas of home
healthcare
10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
Environments in
areas of home
healthcare
83