USER'S MANUAL AE-966 / 967
3.- The RS-232C port configuration appears.
Configure the PC according to this setting.
Baud: 115200
Parity: Ninguno
Stop: 1
Data: 8
4.- Connect the RS-232C cable to the rear
panel terminal.
(9 pins, female)
GPIB (optional)
GPIB interface is a factory installed optional item. Contact the
configuration
service center for a new installation.
1.- Press the System key.
2.- Press F2 (GPIB Add).
3.- Select the GPIB address using the
Left/Right key or Scroll knob. Configure the
PC accordingly.
4.- Connect the GPIB cable to the rear panel
terminal.
Check GPIB
1.- To check the GPIB installation status,
installation
press the System key.
status
2.- Press F6 (More).
3.- Press F4 (System Config).
4.- The system configuration window appears.
GPIB sign shows installed (checked) or not
installed (unchecked)