Check GPIB self-
test result
GPIB constraints
Functionality
Check
Page 158
1.- To check the internal GPIB functionality
test result, press the System key.
2.- Press F6 (More).
3.- Press F2 (Self Test).
4.- The GPIB result appears at F1. If the
result is Fail (underlined), contact the service
center.
Keep these rules when using the GPIB interface.
- Altogether less than 15 devices & 20m cable length, 2m between
each device on the bus.
- Unique address assigned for each device.
- At least 2/3 of the GPIB devices turned On.
- No loop or parallel structure allowed.
Run this query command from the terminal.
*idn?
This should return the Manufacturer, Model number, Serial
number, and Firmware version.
PROMAX, AE-967, P920130, V3.01
USER'S MANUAL AE-966 / 967
System
More...
Self Test...
GPIB
Pass Fail
F 6
F 2
F 1