Instrument Set-Up; Set-Up Location - MACHEREY-NAGEL QUANTOFIX Relax Manual De Instrucciones

Ocultar thumbs Ver también para QUANTOFIX Relax:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Manual QUANTOFIX
Relax
®

3. Instrument set-up

3.1 Set-up location

Only use the instrument in a suitable location. It should be positioned on a dry,
clean, level, hard and horizontal surface. The operating temperature is between
5–40 °C, while the optimal temperature range for the instrument is between
18–25 °C. For optimal performance, the humidity should be between 20–80 %.
Make sure to protect the instrument from drastic changes in temperature. Avoid
positioning the instrument close to windows, hot plates, ovens etc. To ensure
correct functioning and reliable results, the instrument should not be exposed to
direct sunlight or other bright, focused light sources.
If the instrument underwent drastic temperature changes (e.g. during transport
of shipment), it should not be operated prior to sufficient acclimatization. Also,
we recommend not to use the instrument close to electrical fields (e.g. micro
wave ovens, radio units etc.), as it may affect measurements.
57
MN
Manual QUANTOFIX
Relax EN V 1.10 / 01.14
®

Publicidad

Tabla de contenido
loading

Tabla de contenido