12. Check mode
The check mode allows you to check the correct functioning of instru-
ment using special control strips. You can order the control strips from
MACHEREY–NAGEL (REF 913 470).
Once you press
, the instrument asks you to insert the control strip. The
measurement of the control strip starts automatically as soon as the strip is
placed on the slide. Once the measurement has been completed, the measured
values are immediately displayed or printed. The values are so called remission
values for the pre-colored test pads on the control strip. These values have to
be compared to a set of standard values from the control strip instructions. For
additional information on the control strips, please consult the package insert, or
contact MACHEREY-NAGEL directly.
CHECK MODE
Insert
Control strip!
Display 10: Insert control strip