17. Error messages
If you cannot solve a problem with the instructions in the table, please contact
your distributor or MACHEREY-NAGEL directly (see 18.5 Contact, p. 95).
Error messages are displayed as cleartext and are self explanatory.
Error Message
"Remove strip"
"Wrong position"
"No paper"
"Battery empty"
"Instrument doesn't
start"
MN
Manual QUANTOFIX
Manual QUANTOFIX
Source
Test strip has not been
removed from slide af-
ter measurement in the
auto start mode
Strip
has
not
pushed to the far end of
the slide
There is no paper left or
the printer cover is still
open
Empty batteries
No power or power con-
nection broken
Relax EN V 1.10 / 01.14
®
Solution
Remove test strip and
start new measurement
been
Start a new measure-
ment and push the strip
to the end of the slide
Insert new paper and
close printer cover
Change
batteries
connect to an outlet.
Check if all connections
are correctly set and if
the outlet is working.
Relax
®
or
93