By selecting the correction menu four values appear, that can be edited by
choosing each of them. Selecting this function always affects the parameter
displayed in the strip menu (see 13.1 Test strips, p. 79).
The meaning of the values A to D is as follows:
A = gradient of the correction function
B = y-axis intercept of the correction function
C = new lower measurement limit
D = new upper measurement limit
Confirming your entry with
of your selections is adopted. By pressing the "Reset"-button the values will be
set back to factory settings.
Once you select one of these values, the input menu appears and you can enter
a value (Restrictions: The value for A must be > 0. Value D must be bigger than
value C, if values C and D are equal to zero, the factory-provided measuring
range will be adjusted depending upon the entered values for A and B).
82
Relax
®
HCHO
A= 1.000
B= 0.000
C= 0.000
D= 0.000
Reset
Display 16: Correction