Abbott ALERE AFINION AS100 Manual Del Usuario página 29

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 148
When an information code appears
Information codes that might appear during use of the Alere Afinion AS100 Analyzer refer to specific information or error messages.
The code numbers, the possible causes and actions to take are listed below.
If the Analyzer detects a problem during processing of a Test Cartridge, the test will automatically be aborted and the Test Cartridge
will be safely moved to the cartridge chamber. Proceed as follows:
1
Record the code number (#) and touch
The lid opens automatically.
2
Remove the Test Cartridge.
If the Test Cartridge is not ejected, restart the Analyzer.
Do not re-use the Test Cartridge.
3
Look up the possible cause from the table below, and take actions to solve the problem.
If the problem persists, contact your local Alere Afinion supplier (see "Service information" page 29).
Do not re-use a Test Cartridge that has been rejected by the Analyzer. Collect a new sample and repeat the test with a new
Test Cartridge.
Information codes caused by test-specific limitations
[#]
Cause
101
Hematocrit too low
102
Hematocrit too high
103
Hemoglobin too low
104
Hemoglobin too high
105
HbA1c too low
106
HbA1c too high
107
Creatine too high
108
Blood in urine
Information Codes and Troubleshooting
to accept.
Action to take
Consult the Afinion CRP or Lipid Panel Package Insert.
Consult the Afinion CRP or Lipid Panel Package Insert.
Consult the Afinion HbA1c Package Insert.
Consult the Afinion HbA1c Package Insert.
Consult the Afinion HbA1c Package Insert.
Consult the Afinion HbA1c Package Insert.
Consult the Afinion ACR Package Insert.
Consult the Afinion ACR Package Insert.
EN | 27

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido