01/2021
Attention:
Make sure that acid doesn't drip into the reservoir bottle. In case that acid already
dripped into the reservoir bottle, carefully rinse the bottle and the bottle top.
After the cleaning:
Let the apparatus run in distilling mode for one hour. Discard the distillate.
Paul Marienfeld GmbH & Co. KG
Am Wöllerspfad 4
97922 Lauda-Königshofen, Germany
Tax number: 8028008501
VAT reg. no.: DE 811 479 105
Phone: +49 (0) 9343 6272-0
Deutsche Bank Würzburg
Fax: +49 (0) 9343 6272-25
IBAN: DE95 7907 0016 0130 0888 00
info@marienfeld-superior.com
SWIFT: DEUT DE MM 790
www.marienfeld-superior.com
•
As soon as the deposits have
dissolved, drain the acid from the
evaporator via the outlet valve (6)
into the vessel
•
Close the outlet valve (6)
•
Now hang the tubing of the outlet
valve back into the sink.
Rinsing the evaporator
•
Slowly pour approx. one litre of
tap water into the funnel.
•
Carefully fill the evaporator up to
the lower edge of the distillate
drain (1)
•
Now drain the water through the
outlet valve (6) and close the
outlet valve again.
•
Repeat the rinsing three times.
•
Remove the funnel and hang
the overflow tubing into the sink.
Unscrew the protective cap (2) and
keep it on the base of the apparatus
behind the level tube (3).
Commercial Register Mannheim HRA 550306, personally liable
partner: Marienfeld-Verwaltungs-GmbH, 97922 Lauda-Königshofen
Commercial Register Mannheim HRB 560548, Managing Directors:
Gerd Marienfeld, Harry Marienfeld
27