5.9.2 Cleaning of the cooling coil (12)
Attention:
Acetic acid flows through the outlet valve and the overflow connection into the
vessel.
•
As soon as the deposits have dissolved, pour 300 ml of tap water into the
funnel.
•
After the liquid has flown through, hang the tubing of the outlet valve and
the tubing of the overflow connection back into the sink.
Paul Marienfeld GmbH & Co. KG
Am Wöllerspfad 4
97922 Lauda-Königshofen, Germany
Tax number: 8028008501
VAT reg. no.: DE 811 479 105
Phone: +49 (0) 9343 6272-0
Deutsche Bank Würzburg
Fax: +49 (0) 9343 6272-25
IBAN: DE95 7907 0016 0130 0888 00
info@marienfeld-superior.com
SWIFT: DEUT DE MM 790
www.marienfeld-superior.com
•
After the apparatus has cooled
down completely, drain the water
out of the evaporator through the
outlet valve (6). The outlet valve
(6) must remain opened!
•
Hang the tubing ends of the outlet
valve
(6)
and
connection (5) into the vessel for
the disposal of the acid. For safety
reasons place the vessel into a
collection tray.
•
Firmly screw on the protective cap
(2) on the pressure equalization
opening of the level tube (3).
•
Disconnect
the
tubing from the tap and insert the
funnel into the free end of this
tubing. Make sure that the funnel
fits tightly!
•
Now insert the funnel in the stand.
The lower end of the funnel cone
must be on the same level as the
degassing outlet (1).
•
Use the graduated jug to pour
approx. 300 ml of acetic acid (10
to 50 %) into the funnel. Be
carefully and avoid spilling!