Figure 3-22 User Maintain Menu
• LANGUAGE :language choice.
• LEAD NAMING:AHA/EURO
• ALM SOUND:ON/OFF
• HELP SETUP:ON/OFF
• NIBP OBSTRUCT SETUP: 1/2/3/4
This function is used to detect whether the patient moves during the blood pressure measurement. If the pa-
tient moves, the monitor will give an alarm message and stop the current measuring, or the measurement
will be taken as usual.
1) This function is set "1" as default.
2) "1" represents the sensitivity is reduced to the minimum, "4" represents the sensitivity is increased to the
maximum. The higher level the sensitivity is set and the easier to detect the interference of movement.
• "NETWORK CONFIGURATION": see Section 3.7.1 Network Configuration for details.
HL7 server configuration:
IP: 202.114.4.120. Input the IP address of the server.
Port: 511. Input the server port.
Sending interval: 1. Set the frequency of data sending, unit is "second".
• ALARM SET PASSWORD MODIFICATION: it is used to change the login password of "ALARM SETUP".
Warning
• When the alarm volume is set to "OFF", you will not hear the alarm sound if new alarm occurs.
Therefore, you must be very careful in using this selection.
• If setting the alarm volume to "OFF" when the system is in Silence or Pause status, the system will
automatically discharge Silence or Pause status.
• If you select "Silence" or "Pause" when the alarm volume is set to "OFF", the system will restore the alarm
volume before the alarm volume is set to "OFF" and enter Silence or Pause status.
Note
• After the alarm volume is set to OFF, a symbol will appear in the Technical Alarm Area.
• Setting Alarm Volume to "OFF" is valid only when the monitor is turned on for this time. After turning on the
monitor next time, this setup will restore its value of the previous time when the system is turned on.
3.7.1 net ConFIG
Press "NET CONFIG" to pop the following menu:
Figure 3-23 NET CONFIG
• NET TYPE:CMS/CUSTOM
CMS: the Server IP is fixed, "202.114.4.119", "LOCAL IP CONFIG" is unavailable.
CUSTOM: when this item is selected, CMS and machine's IP can be changed as you need.The following is
"LOCAL IP SETUP" menu.
24