• ARR ANAL:Pick "ON" during monitoring. Default set is "OFF".
• PVCs ALM: pick "ON" to enable prompt message and data record when alarm occurs; pick "OFF"
to disable the alarm function, and there will be a beside "PVCs".
• ALM LEV: selectable from HI, MED, LO. Level HIGH represents the most serious case.
• ALM REC: pick "ON" to enable report printing upon PVCs alarm.
PVCs alarm is activated when the PVCs exceeds set PVCs ALM HI value.
PVCs alarm upper limits:
Max
PVCs
30
PVCs alarm and prompt message:
Among physiological alarms, those belonging to the type that the parameter has exceeded the limits may
activate the recorder to automatically output the parameters and related measured waveforms when the
alarms occur on the condition that the alarm record switch in the related menu is On.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring
during pvcs measurement.
Physiological alarms:
Message
Cause
PVCs TOO HIGH PVCs measuring value is above upper alarm limit.
Technical alarms:
Message
PVCs ALM LMT ERR
• ARR RELEARN Pick this item to start a learning procedure.(unusable)
• ARR ALARM
Pick this item to access the ARR ALARM dialog box to set arrhythmia alarm parameters.
Set ALM to ON/OFF to enable/disable the alarm function; Set REC to ON/OFF to enable/disable alarm
record function, turn the knob under LEV column to set alarm level to HIGH, MED or LOW.
Figure 11-12 ARR Alarm Menu
You can pick ALL ALM ON to enable alarm function of all arrhythmia types and pick ALL ALM OFF to disable
this function. Likewise, you can pick ALL REC ON to enable recording function for all arrhythmia types and
pick ALL REC OFF to disable this function. Changing the ALM LEV can reset alarm level of all arrhythmia
types to the same value.
• ARR RECALL Pick this item to review and edit the ARR analysis result.
The latest arrhythmia events (up to 60) are displayed.
Min
1
Cause
Functional safety failure
Step
1
Alarm Level
Remedy
HIGH
Stop using PVCs alarming
function, notify biomedical
engineer or Our service staff.
Alarm Level
User-selectable
60