Alarm information:
SpO
NO SENSOR
2
SpO
SENSOR OFF
2
SpO
SENSOR FAULT
2
SpO
UNRECOGNI-
2
ZED SENSOR
SpO
INCOMPATIBLE
2
SENSOR
SpO
INTERFERENCE
2
SpO
PULSE
2
SEARCH
SpO
LOW
2
PERFUSTION
SpO
TOO MUCH
2
LIGHT
SpO
LOW SIGNAL
2
IQ
SpO
BOARD FAULT
2
SpO
2
COMMUNICATION
ERROR
SpO
2
COMMUNICATION
STOP
SpO
INIT ERR
2
"TEMP1 SENSOR
OFF"
"TEMP2 SENSOR
OFF"
"IBP1 LEAD OFF"
"IBP2 LEAD OFF"
Sensor not fully inserted into
the connector.
Sensor inserted upside down.
SpO2 sensor may be disconnected
from the patient or the monitor.
This message appears when the sen-
sor is faulty
board does not recognize the sensor.
This message is displayed when the
sensor is finding incompatible sensor.
Outside signal or energy preventing
reading.
Unit is searching for the patients
pulse.
Signal too small.
Too much light on patient(sensor).
Inadequate tissue covering sensor
detector.
Low signal quality.
This message appears when the Set
board malfunctions.
This message is displayed when the
front end module is having problems
communicating ( ie: framing errors
or bad checksums) with the board.
This message is displayed when
the host can not receive the data from
board for 5 seconds
This message is displayed when
the SpO2 module initialization error
happened.
TEMP1 sensor is not connected
correctly.
TEMP2 sensor is not connected
correctly.
IBP1 sensor is not connected
correctly.
IBP2 sensor is not connected
correctly.
May be an incorrect sensor, or a
defective sensor or cable. Insert sensor
into the connector. Disconnect and
reconnect sensor. Refer to the
instructions for the sensor being used.
Disconnect and reconnect he sensor
with the logos matching.
Disconnect and reconnect the sensor.
Reattach sensor.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Make sure that the monitor and the
patient are in correct connection with
the cables.
Make sure that the monitor use
incompatible sensor.
Remove outside interference.
If values are not displayed within 30
seconds, disconnect and reconnect
sensor. If pulse search continues, remove
sensor and replace on a better perfused
site.
Move sensor to better perfused site.
Remove or reduce lighting. Cover sensor
from light.
Reposition sensor.
Ensure proper sensor application. Mover
sensor to a better perfused site.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Stop using the measuring function of
SpO2 module, notify biomedical engineer
or our service staff.
Check the connection of TEMP1 sensor.
Check the connection of TEMP2 sensor.
Check the connection of IBP1 sensor.
Check the connection of IBP2 sensor.