109
"5V TOO HIGH"
"5V TOO LOW"
"POWER ERR3"
"POWER ERR4"
"12V TOO HIGH"
"12V TOO LOW"
"POWER ERR7"
"POWER ERR8"
"3.3V TOO HIGH"
"3.3V TOO LOW"
"CELL BAT TOO
HIGH"
"CELL BAT TOO LOW"
"RECORDER
SELFTEST ERR"
"RECORDER VLT
HIGH"
"RECORDER VLT
LOW"
"RECORDER HEAD
HOT"
"REC HEAD IN
WRONG POSITION"
"RECORDER OUT OF
PAPER"
"RECORDER PAPER
JAM"
"RECORDER COMM
ERR"
"RECORDER S.
COMM ERR"
"RECORDER PAPER
W.P."
"REC NOT
AVAILABLE"
"NIBP INIT ERR"
"NIBP SELFTEST
ERR"
The power part of the system has
failure.
Cell battery has problem.
The cell battery has low capacity or
the cell battery is not installed or the
connection is loose.
During the selftest, the system fails
connecting with the recorder module.
The recorder module has voltage
failure.
The continuous recording time may be
too long.
The handle for pressing the paper is
not pressed down.
No paper is in the recorder.
The paper in the recorder is jammed.
The communication of the recorder
is abnormal.
The paper roll of the recorder is not
placed in the correction position.
Cannot communicate with
the recorder.
NIBP initialization error
If the prompt appears repeatedly, contact
the manufacturer for repair.
Replace the battery. If the failure still
exists, contact the manufacturer.
Execute 'Clear Record Task' function
in the recorder setup menu to re-connect
the host and the recorder. If the failure
still exists, contact the manufacturer
for repair.
Contact the manufacturer for repair.
After the recorder becomes cool, use
the recorder for output again. If the failure
still exists, contact the manufacturer
for repair.
Press down the recorder handle for
pressing the paper.
Place the paper into the recorder.
Place the recorder correctly and try again.
In the recorder setup menu, execute
the function of clearing record task.
The function can make the host and
the recorder connect again. If the failure
still exists, contact the manufacturer
for repair.
Place the paper roll in the correct
position.
In the recorder setup menu, execute
the function of clearing record task.
The function can make the host and
the recorder connect again. If the failure
still exists, contact the manufacturer
for repair.
Execute the reset program in the NIBP
menu. If the failure still exists, contact the