Physiological alarms:
Message
Cause
IS1 TOO HIGH
SYS measuring value of channel 1 is above
upper alarm limit.
IS1 TOO LOW
SYS measuring value of channel 1 is below
lower alarm limit.
ID1 TOO HIGH DIA measuring value of channel 1 is above
upper alarm limit.
ID1 TOO LOW
DIA measuring value of channel 1 is below
lower alarm limit.
IM1 TOO HIGH MAP measuring value of channel 1 is above
upper alarm limit.
IM1 TOO LOW
MAP measuring value of channel 1 is below
lower alarm limit.
IS2 TOO HIGH
SYS measuring value of channel 2 is above
upper alarm limit.
IS2 TOO LOW
SYS measuring value of channel 2 is below
lower alarm limit.
ID2 TOO HIGH DIA measuring value of channel 2 is above
upper alarm limit.
ID2 TOO LOW
DIA measuring value of channel 2 is below
lower alarm limit.
IM2 TOO HIGH MAP measuring value of channel 2 is above
upper alarm limit.
IM2 TOO LOW
MAP measuring value of channel 2 is below
lower alarm limit.
Technical alarms:
Message
IBP1 SENSOR OFF
IBP2 SENSOR OFF
IBP (1,2) INIT ERR
IBP (1,2) INIT ERR1
IBP (1,2) INIT ERR2
IBP (1,2) INIT ERR3
IBP (1,2) INIT ERR4
IBP (1,2) INIT ERR5
IBP (1,2) INIT ERR6
IBP (1,2) INIT ERR7
IBP (1,2) INIT ERR8
IBP (1,2) COMM STOP
IBP (1,2) COMM ERR
Cause
IBP cable of channel 1 falis
off from monitor.
IBP cable of channel 2 fails
off from monitor.
IBP module failure
IBP(1,2) module failure or
communication failure
IBP(1,2) communication
error
Alarm Level
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
Alarm
Remedy
Level
LOW
Make sure that cable is properly con-
nected.
LOW
Make sure that cable is properly con-
nected.
Stop using measuring function of IBP
HIGH
module, notify biomedical engineer or
Our service staff.
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
86