TECHNICAL DATA
DESCRIPTION
Manufacturer
Model
Classification 93 / 42 EEC
Mode
Service Pressure to the turbine con-
nection : Water
Service Pressure to the turbine con-
nection : Air
Weight
Operating conditions
Storage and transport conditions
EMS SA, CH-1260 Nyon, Switzerland
AIR-FLOW
handy 2
handy 2
®
®
®
+
Class IIa
Continuous operation
0.7 to 2.2 bar (700-2200 hPa)
with a service flow of 50-80 ml/min.
Static pressure 3.5 to 4.5 bar
(3500-4500 hPa)
Approx. 0.160 kg
+10° C to +40° C
30% to 75% relative humidity
500 hPa to 1060 hPa air pressure
-10°C to +40°C
10% to 95% relative humidity
500 hPa to 1060 hPa air pressure
CARACTÉRISTIQUES TECHNIQUES
DESCRIPTION
Fabricant
Modèle
Classification 93 / 42 CEE
Mode de fonctionnement
Pression de service au raccord de
turbine: Eau
Pression de service au raccord de
turbine: Air
Poids
Conditions de fonctionnement
Conditions de stockage et de trans-
port
EMS SA, CH-1260 Nyon, Suisse
AIR-FLOW
®
®
®
handy 2
handy 2
+
Classe IIa
Service continu
0.7 à 2.2 bar (700-2200 hPa)
avec un débit de 50-80 ml/min
Pression statique de 3.5 à 4.5 bar
(3500-4500 hPa).
Approx. 0.160 kg
+10°C à +40°C
30% a 75% d'humidité relative
pression atm. 500 hPa à 1060 hPa
-10°C à +40°C
10% à 95% d'humidité relative
pression atm. 500 hPa à 1060 hPa
59