TROUBLESHOOTING
Type of problem
Water comes into the powder chamber
or leaks from the cap
No powder/air jet coming from the
unit
Air and/or powder is leaking from the
threads of the cap
The unit efficiency diminishes
62
Solutions
•
Check the dental unit connection
•
Check the state of the o-rings of the turbine connection
•
Clean the cap and the powder chamber
•
Clean the handpiece
•
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
•
Check the quality of the spray
•
Fill the powder chamber
•
Send the unit to an approved EMS repair center
•
Cut immediately the air input relensing the foot pedal
•
Wait for 1-2 minutes for the system to decompress
•
Disconnect the unit from the dental unit
•
Press the foot control pedal of the dental unit
•
If no air comes out for the turbine connector, it is your dental unit that causes the problem
•
If air comes out, it is the device that causes the problem
•
Unscrew the cap with the top of the wash-hand basin, bus of the powder can be ejected. A sealed appara-
tus, even disconnected, can indeed remain under pressure
•
Empty the powder chamber and screw the cap back on
•
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
•
Connect the handpiece to the unit
•
Press the foot control pedal of the dental unit
•
If air comes through the handpiece connector, the handpiece is blocked. Clean the handpiece
•
If no air comes out of the handpiece connector, the unit is blocked
•
Send the unit to an approved EMS repair center
•
Check the seal and the cleanliness of the thread on the powder chamber and on the cap
•
Replace the seal if necessary
•
A powder refill is perhaps necessary
•
Clean the handpiece
•
Send the unit to an approved EMS repair center