Maintenance And Care; Care Of The Handpiece; Reprocessing The Light Guide - 3M Elipar DeepCure-L Instrucciones De Uso

Ocultar thumbs Ver también para Elipar DeepCure-L:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 59
SEITE 9 - 148 x 210 mm - 3100021788/01 - SCHWARZ - 20-052 (kn)
Error
Cause
While the handpiece is
The handpiece is connected
connected to the
to the charger. Light-curing
charger, pressing the
is prevented as a safety
START button will not
precaution.
start the light emission.
The light emission
The information about inter-
does not start when
mittent operation under the
the START button is
header "Technical Data -
pressed; an error
Handpiece" has not been
signal sounds for
observed.
2 sec.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
During light emission
The information about inter-
in continuous mode,
mittent operation under the
an error signal sounds
header "Technical Data -
for 2 sec, the emission
Handpiece" has not been
is stopped, and the
observed.
handpiece changes
The handpiece has become
to sleep mode.
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
The light intensity is
too low.
Solution
Disconnect the charger
from the handpiece and
restart the light emission.
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
Clean the light guide and
the protecting glass in the
light guide mounting hole
(please refer to "Cleaning
the Light Guide").
Error
Cause
The dental material
does not cure
completely.
The light guide cannot
be attached to the
handpiece.

Maintenance and Care

The Elipar DeepCure-L device is maintenance-free.
No periodic maintenance is required. See the informa-
tion contained in this chapter to secure problem-free
operation.

Care of the Handpiece

Use only the charger included with the product. The
use of other chargers may damage the battery cells
or result in inadequate charge.
Do not immerse the handpiece in water or incinerate.
Please also observe the chapter on "Safety".

Reprocessing the Light Guide

The light guide is not sterile when delivered and must
be reprocessed before being used for the first time and
after each patient use.
Follow manufacturer's instructions for preparation, tem-
peratures, contact time and rinsing off the cleaning
and disinfecting solution
2
exactly.
An enzymatic detergent should be used as the cleaning
solution.
Material Resistance
Make sure that the cleaning and disinfectant agents you
have chosen do not contain any of the following materi-
als:
– Organic, mineral, and oxidizing acids (minimum ac-
ceptable pH value 5.5)
– The pH value of the cleaning and disinfecting solu-
tion should be between pH 7 and 11
– Oxidation agents (e.g., hydrogen peroxide)
– Halogens (chlorine, iodine, bromide)
– Aromatic/halogenized hydrocarbons
Solution
Clean the light guide and
the protecting glass in the
light guide mounting hole
(please refer to "Cleaning
the Light Guide").
Check to see that the
correct light guide has
been attached.
The light guide is not
designed for use with the
Elipar DeepCure-L.
1
9

Publicidad

Tabla de contenido
loading

Tabla de contenido