Section 5
To modify the program:
Go to [Programming], make adjustments, then save it on the Remote
Control.
Preview
Before overwriting the currently stimulating program with a program from the Remote
Control, a program in the Remote Control can be previewed. Click on [Preview]. The
currently stimulating program will be stopped and the selected program will be activated
temporarily until [End Preview] is clicked. When [End Preview] is clicked, the original
program is activated. This feature can help in determining which Remote Control
program to overwrite when all program slots are filled.
Click the desired [Program slot] from the Remote Control then select [Preview].
Click, [End Preview] to return to the previous active program.
Delete
Click on a [Program slot].
To erase the program from the Remote Control select [Delete].
Additional Information
Click [Patient Options] to change lockouts.
Click [VAS] to document pain levels.

Stimulation Power Button

Description
The Stimulation (Stim) Power button will stop and start stimulation. When started,
stimulation will ramp up to the stored amplitude levels for the active program settings.
Operation
Click the red [Stop] button in the lower left-hand corner of any screen to stop
stimulation.
Click the green [Turn on] button to resume stimulation.
Programming and Patient Fitting
Bionic Navigator HR 1.0 Software Guide
90868740-07 Rev A 9 of 84