Bionic Navigator
HR 1.0 Software Guide
Section 1
Clinician Programmer

Text Entry

Description
Text entry options are provided for ease of data entry. There are two methods for
entering patient Profile data and patient Notes using the CP keyboard and Tablet PC
Input Panel feature. The text entry icon for the Tablet PC Input feature may be located
on the Windows Task Bar.
Access/Exit
To access the [Windows Task Bar], move the pen over the very bottom of the
screen. Click the keypad [Icon].
To exit any text entry tool, click the [X] on the right-hand corner of the tool.
CP Keyboard
The CP Keyboard is used to enter patient information using the conventional method.
Tablet PC Input Panel
The Tablet PC Input Panel is a touchable keyboard and writable window.
Place the cursor on the [Text field]. Enter the appropriate letters or numbers on
the keypad or write the text in the text input window.
To erase information, click Back Space ( < –).
Additional Information
See Toshiba Portege M400 Tablet PC Series Manuals at www.toshiba.com
for additional instructions.
See the Set-up and Communication section below for more information.

Set-Up and Communication

Description
The Clinician's Programmer (CP) communicates with the stimulator using an infrared
link to the Remote Control. Programs can be loaded to the CP for previewing,
activating and changing stimulation patterns. Stimulation programs can also be saved to
the Remote Control.
Bionic Navigator HR 1.0 Software Guide
90868740-07 Rev A 2 of 84
Set-up and Operation
Caution: The CP is not equipment for the patient environment as defined by
IEC 60601- 1. The CP and the person using the CP should not be in contact
with the patient while programming.
Power [ON] the CP to display the Neuromodulation desktop.
Select the Bionic Navigator™ icon
1.0 Software.
1. Enter the [Clinician's screen] on the Remote Control by pressing both the
P
Area
and Program
buttons simultaneously.
"
2. Select "CP Mode".
When the Remote Control displays [CP Ready], face the IR window at the top of
the Remote Control towards the IR extension.
Position the patient within two feet of the Remote Control to ensure
communication with the implantable stimulator.
Maintain a clear path between the IR extension attached to the Clinician's
Programmer and the Remote Control IR window.
Using the Patient Controller (Joystick)
Adjust the stimulation level by pressing the up and down arrow buttons.
To stop or turn off stimulation immediately, press the [OFF] button.
Additional information
A message will be displayed on the Clinician Programmer if the communication
range is not satisfactory.

System

Description
The system screen provides access to selecting stimulator maximum rate, backing up
Database and Logs, estimating relative lead positions, and stimulator information.
Access/Exit
To access "System," select [System] at the top right-hand corner of the screen,
make adjustments.
To exit, click [Close].
Stimulator Maximum Rate
Click [Set Stimulator Maximum Rate] to select the maximum rate at which the
connected stimulator can operate. Note that changing the maximum rate of the
connected stimulator will clear all settings and programs from that stimulator.
to boot up the Bionic Navigator HR