English
Temperature
Relative Humidity
Atmospheric Pressure
Primary: 100-240VAC, 50/60 Hz, 250 VA
Class I: Protective Earth
Fluid Ingress Protection: IPX0
Type B: Applied Part, Continuous Operation
Note: If a fuse replacement is necessary, use only a Time Lag, 5 x 20 mm, 5A, 250V fuse. Item to be replaced by
DePuy Synthes Power Tools repair facility ONLY. Return all equipment to DePuy Synthes Power Tools for proper
disposal.
Caution: For continued protection against risk of fire and electric shock, replace fuse only as marked.
Guidance and manufacturer's declaration –electromagnetic emissions
The eG1 System is intended for use in the electromagnetic environment specified below. The customer or the
user of the eG1 System should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
12
Environmental Conditions
Operating: 18°-30°C (65°-85°F)
Transportation and Storage: -29° – +50°C (-20° - +120°F)
Operating: 30% to 70%
Transportation and Storage: 10% to 90%
Operating: 0.7–1.06bar
Transportation and Storage: Not applicable
Console (SC3001 and SC3002) Specifications
Compliance
Electromagnetic environment – guidance
The eG1 System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
Group 1
not likely to cause any interference in nearby electronic
equipment.
Class A
The eG1 System is suitable for use in all establishments
Class A
other than domestic establishments and those directly
connected to the public low-voltage power supply network
which supplies buildings used for domestic purposes.
Complies