1. Pre - Installation Checks

Prior to fitting the service drive unit the concentricity of the AVM2-PC anti-backlash coupling fitted to the
test stand must be checked and it must not exceed the limits as outlined below.
Max Limit 0.008" / 0.20MM TIR
To check the concentricity of the AVM2-PC anti backlash coupling the following procedure should be
followed: -
Equipment required: - Magnetic base and DTI (dial indicator with a resolution 0.001"/0.2mm).
1. Fit drive guide 182-18-17 to the anti backlash coupling (originally supplied with the AVM2-PC).
2. Locate the magnetic base and the DTI onto the AVM2-PC test bed next to the AVM2-PC anti
backlash coupling.
3. Position the probe of the DTI indicator on the AVM2-PC anti backlash coupling as shown in the
diagram below (fig 3) so that the indicator can move up and down and set to zero.
DTI TIP
4. Rotate AVM2-PC anti backlash coupling 1 complete turn using the AVM2-PC Test stand nudge
function of the AVM2-PC. Monitor the DTI indicator during the revolution. There should be no more
than 0.008" / 0.20MM variation.
5. If the concentricity of the AVM2-PC anti backlash coupling is greater than specified above, do not
attempt to fit or run the ADC298 Lubricated Service Drive Unit. Contact your distributor immediately
for advice.
8
THIS IS AN UNCONTROLLED DOCUMENT downloaded by Lukas Matuska on 16 Feb 2016
Any technical intervention requires certified Hartridge training. Contact Hartridge Ltd for details.
ROTATE DRIVE USING THE NUDGE FUCTION, DO
NOT ROTATE BY HAND
MAGNETIC BASE & DTI SHOULD BE FITTED TO
THE BED OF THE MACHINE
HARTRIDGE LIMITED
Fig 3
MaualRefNo HL033., IssueNo. 1