7.2.3 - Saliva Ejectors and Solid's Collector
The Saliva Ejectors and Solid's Collector must also be cleaned
daily. The suction power may be decrease if this collector is blocked. In
case of reduction in the suction power, clean the collectors.
To clean the solid collector, follow the instructions below:
1º - Undo the solid collector cover (1);
2º - Remove the solid collector (2) for cleaning;
3º - After cleaning, re-assemble the solid collector.
7.2.4 - Water Unit and Strainer
To clean the spittoon bowl (1),
remove the spittoon bowl's finishing (5),
the strainer (4), and the detachable water
spouts (2 and 3), for then remove the
spittoon bowl.
When the water unit has a steady
spittoon bowl's water spout (6) pay
attention to the correct way to remove the
spittoon bowl.
For removal of the strainer (4) use a
tweezer or gloves to avoid direct contact
with the waste.
The
cleaning
strainer (4) and the spittoon bowl (1) can
be made with running water and neutral
soap or detergent. Use a non-abrasive sponge.
7.3 - Sterilization by Autoclave
Prior to autoclaving, clean the items, removing all organic residue, both from the surface and internal
ducts (if any). Then carefully dry each item, including the internal ducts, if possible, with compressed air.
Do not use any type of oil on the items to perform autoclaving.
Individually pack each item, with its own sterilized packaging for the autoclaving process.
For sterilization use steam autoclave: 134º C; 2.2 bar; 4 minutes.
Note: the scaler transducer with LED cannot be autoclaved in contact with other materials. The same applies to
the bicarbonate jet handpiece and electrical micromotor Brushless.
Regarding the resistance to the autoclaving process, the following items support:
− Up to 1000 cycles: scaler tip, stainless steel tray, 3-way syringe tip, optical fiber curing light tip and electrical
micromotor Brushless;
− Up to 600 cycles: transducer and LED of the scaler, bicarbonate jet handpiece, and Prophy-Jet removable
tip.
7.4 - Curing Light
To clean and disinfect the curing light tips, use gauze or disposable wipe moistened with soap and neutral
detergent. Do not use alcohol, strong or abrasive alkaline detergents, detergents based on bleach, acetone, or
other germicides.
The whitening tip cannot be autoclaved.
Curing light's body cleaning must be done with neutral detergent or 70% alcohol.
Olsen is not responsible for defects, deformities, spots on or abnormalities caused by improper use
of chemical products, contact with tissues, leather, disposable gloves, inks, pigmented detergents
and other organic or synthetic products.