7.5 - Scaler

7.5.1 - Scaler with LED

The transducer (5), the LED (3), the tips and the scaler
wrenches (6) must be autoclaved for sterilization in
accordance with the instructions in section 7.3 (Sterilization by
Autoclave).
Prior to the transducer sterilization, remove the tip of the
scaler (if it is attached), LED removable cover (1), the light
driver (2) and the finishing ring (4). Clean these items by
removing visible dirt with a soft sponge or cloth and running
water. For cleaning in an ultrasonic tub, use appropriate pure
water and a minimum of 2 minutes and a maximum of 10
minutes.
For disinfection use a 75% alcohol or 2% glutaraldehyde solution to
disinfect the tips. Dry the tips at the end of the procedure with a hot air blower.
For the proper functioning of the scaler, the LED (3) must be correctly
inserted in the transducer (5) matching the LED's positive pole with the
transducer's positive pole.
Only use dental autoclave for sterilization of tips, wrenches, transducer (5) and LED (3).

7.5.2 - Standard Scaler

The standard scaler (7) can be cleaned with any neutral liquid
for cleaning and sterilization. Do not autoclave.
For both standard scaler or LED, the following sterilization
methods are forbidden:
− Put in boiling water;
− Heat in stove, oven or microwave;
− Soak in disinfectants such as iodine, alcohol or glutaraldehyde.

7.6 - Daily Procedure at Working Day

At the end of the working day, observe the following instructions:
− Provide the equipment cleaning by cleaning the ejectors and their hoses, the water unit (spittoon bowl),
the upholstery and if necessary clean the operating light screen, the plastic covers, and metal parts;
− Remove the handpieces that were used during the day (high and low-speed) and provide their lubrication.
After the lubrication, sterilize them;
− Remove the other handpieces used during the day: prophylaxis handpiece, 3-way syringe tip, the
transducer, and its tips, curing light tips and provide the sterilization of each one;
− Turn off the compressor circuit breaker and open the air reservoir drain;
− Depressurize tanks, remove water and clean filters;
− Remove the bicarbonate of the prophylaxis reservoir;
− Close the water and compressed air main valves that supply the equipment;
− Run the Zero Position command and turn the equipment On/Off switch Off;
− Organic waste, contaminated and disposable materials, must be disposed of properly as determined by
the responsible local health agency;
− Follow the instructions in chapter 7 (Cleaning and Disinfection) for equipment proper cleaning and its parts.