Meter and Lancing Device Cleaning and Disinfecting
How to Clean and Disinfect the Meter
WARNING
Failure to follow these instructions will damage the meter and stop it
from working properly.
• DO NOT clean or disinfect the meter while performing a blood glucose
or control test.
• DO NOT get any moisture in slots or openings.
• DO NOT spray anything onto the meter.
• DO NOT immerse the meter in liquid.
• Always use the same product for both cleaning and disinfecting.
1
Wash hands with soap and
water and dry thoroughly.
2
Turn the meter off and wipe the
entire meter surface with a
Super Sani-Cloth. Carefully
wipe around the test strip slot
and other openings.
Make sure that no liquid enters
any slot or opening.
3
A separate Super Sani-Cloth
should be used for cleaning and
disinfection. For disinfecting the
meter, get a new cloth and
repeat step 2, making sure the
surface stays wet for 2 minutes.
Make sure that no solution is
seen in any slot or opening.