IEC 61326-1: 2013, IEC 61326-2-6:2013
Test Condition
Radiated Emission
(30MHz to 1000MHz)
Conducted Emission on AC 150kHz to 30MHz
Harmonic Class A
Flicker
IEC 61326-1: 2013, IEC 61326-2-6:2013
Immunity Tests
Electrostatic Discharge (IEC 61000-4-2) ±4kV ±8kV
Radiated Immunity (IEC 61000-4-3) 80MHz to 1000MHz
3V/m (rms) 1400MHz to 2000MHz 3V/m (rms)2000MHz to
2700MHz 1V/m (rms)
Electrical Fast Transient (IEC 61000-4-4) ±1kV
Surge (IEC 61000-4-5) ±0.5kV (Line to line) ±1kV (Line to
ground)
Conducted Immunity (IEC 61000-4-6) 150KHz to 80MHz
3Vrms
Voltage Dips and Interruption (IEC 61000-4-11) 0%, 70% of Ut Complies
FCC Interference Statement
Federal Communication Commission Interference Statement
This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 18 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential
installation. This equipment complies with FCC radiation
exposure limits set forth for an uncontrolled environment when
used with Telcare, LLC accessories supplied or designated
for this product. Use of other accessories may not ensure
compliance with FCC RF exposure guidelines. However, there
is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference
to radio or television reception, which can be determined by
Compliance
Complies
Complies
Complies
Complies
Compliance
Complies
Complies
Complies
Complies
Complies