Priming The Administration Set
Use Aseptic Technique
Note: In order to avoid the risk of precipitation within the
tubing, the HOMEPUMP ECLIPSE* Pump can be prefilled and
primed with a small amount of diluent so that the medication
is not in the pump tubing until the infusion is started. This
technique can be used for any medication prone to precipitation.
Caution: It is important to completely prime the pump
tubing. Failure to do so may prevent the pump from infusing.
1. Remove distal end cap.
2. Open clamp to start priming.
3. When all air has been removed from the entire tubing
and fluid flow is observed at end of the distal luer, the
administration set is primed.
4. Close clamp and replace distal end cap until ready for use.
Starting Infusion
Use Aseptic Technique
Note: Patient must be educated on proper use of product by
healthcare provider.
1. Allow the HOMEPUMP ECLIPSE* Pump to reach room
temperature before using. (Table 1)
2. Verify that the clamp on the tubing is closed.
3. Cleanse patient's catheter injection site/IV access device.
4. Attach the HOMEPUMP ECLIPSE* pump tubing to the
access device.
5. Begin infusion by opening the clamp; fluid delivery will
start immediately.
• Note: If tubing is kinked, roll kinked portion of the tubing
between fingers to restore shape of tubing and promote
fluid flow.
End Of Infusion:
• Infusion is complete when the elastomeric membrane is
no longer expanded.
• Close clamp, disconnect and dispose of the HOMEPUMP
ECLIPSE* pump according to your institution's protocol.
Note: If the pump did not perform as expected do not discard
pump. Contact Halyard Health for product return instructions:
www.halyardirvineproductcomplaint@hyh.com.
Technical Specifications
Delivery Accuracy: When filled to the labeled (nominal)
volume, the HOMEPUMP ECLIPSE* Pump is ±15% of the
labeled (nominal) flow rate when infusion is started 4 hours
after fill , delivering normal saline as the diluent at 20°C/68°F.
Typical Flow Curve
Below is a graph of the typical flow profile of the HOMEPUMP
ECLIPSE* Elastomeric Pump. The flow rate may be higher or
lower than ±15% at the start and end of infusion.
Typical Flow Rate
Percent Delivery Time
Notes:
• Latex is not in fluid pathway or in contact with human.
Refer to Latex Sensitivity Technical Bulletin at
www.halyardhealth.com.
• Product is NOT made with DEHP (Di(2-ethylhexyl)phthalate)
as a plasticizer.
• Length of tubing is approximately 96 ±20 cm (38 ±8 inches)
Storage Conditions
Store under general warehouse conditions. Protect from light
sources and heat. Keep dry.
Nominal
100%
5