Guidance and manufacturer's declaration

EMC compliance

Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical
electrical equipment. This device complies with IEC EN 60601-1-2:2007.
All medical electrical equipment must be installed and put into service in accordance with the
EMC information provided in these tables and in the Instructions for use.
Portable and mobile RF communications equipment can affect the behavior of medical
electrical equipment.
The MacroView and LED Otoscopes and PanOptic Ophthalmoscopes devices comply with all
applicable and required standards for electromagnetic interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is not safe to operate the device in the presence of high-frequency surgical equipment.
However, it is good practice to avoid using the device in extremely close proximity to other
equipment.
NOTE The device has essential performance requirements associated with Welch
Allyn physical assessment handles 710, 716, 719, 767, 777, and CIWS. In the
presence of EM disturbances, the device may not perform as intended. Once the
EM disturbances stop, the device self-recovers and performs as intended.
WARNING Use only accessories and cables Welch Allyn recommends for use with
the device. Accessories not recommended by Welch Allyn may affect the EMC
emissions or immunity.
WARNING Maintain minimum separation distance of 12 inches (30 cm) between
any part of the device and portable RF communication equipment (including
peripherals such as antenna cables and external antennas). Performance of the
device might degrade if proper distance is not maintained.
WARNING The use of the device adjacent to or stacked with other equipment or
medical electrical systems should be avoided because it could result in improper
operation. If such use is necessary, the device and other equipment should be
observed to verify that they are operating normally.