Note: The temp command enables the heating of a sample to 37 °C during a series of program steps, e.g., for
incubation purposes. The incubation period will only include the steps between a temp ON and temp OFF
command.
11. Highlight the end command and press Add. In this example, two spin commands are added, a counter-
clockwise spin at 20 rpm for 30 minutes and a clockwise spin at 3,000 rpm for 10 seconds.
4
12. Highlight the end command and press Add. Select temp and OFF from the pop-up screen to switch off the
heating function. Press OK to save the value.
13. Press Save to store the new program. Enter the filename when prompted. In this example, the new program is
named program_03.
70