TROUBLESHOOTING
The thermostat does not switch on -
 Make sure the device is properly connected to the power supply.
 Check that the power cord is intact. Do not use the device if the power cord is damaged.
 Make sure that the power switch is set to ON. The LCD screen should also switch on when
the device is switched on.
The LCD displays the error code E01 -
 Make sure that the water level in the tank is between the minimum and maximum values
indicated on the thermostat. The device will turn off and give a special sound alarm if the water
level drops below the minimum mark.
The LCD displays the error code E05 -
 Thermostat pump is defective. The pump needs to be replaced. Contact the service center.
The LCD displays the error code E06 -
 Thermostat fan is defective. The fan needs to be replaced. Contact the service center.
The device is plugged in and switched on, but the LCD does not display any information -
 Remove the plastic cover from the front of the device and press the RESET button.