8.
ADVANCED SETTINGS [e-IFU]
These settings are only intended to be set by qualified caregivers or qualified technicians.
9.
ALARMS, WARNING MESSAGES AND TROUBLE SHOOTING [e-IFU]
The Flocare Infinity III is equipped with several sensors and safety features linked to clear visual and audio alarms and warning messages on the display.
In case any of the alarms occur, the pump will generate an audible and visual alarm and stops administering. The back light of the display automatically switches on.
Exception to this is the low battery alarm, in this situation the pump continues working.
If an alarm or warning message appears on the display:
The HOME [
] button acknowledges an alarm or warning state and dismisses the alarm or warning message.
The SNOOZE [
] button snoozes the alarm or warning audio signal for 2 minutes whilst the alarm or warning message remains on the display. A
suppressed audio icon will appear on the display indicating that the audio signal has been snoozed.
The INFORMATION button [
Note: during an alarm or warning, all other buttons are disabled. The HOME [
message.
10.
CLEANING, INSPECTION & SERVICING GUIDELINES [e-IFU]
For optimum and safe pump operation, the pump should be kept clean and free from debris. It is recommended to clean the pump with a wet cloth after every
completed feeding regimen.
It is recommended to clean the pump regularly under (running) warm tap water with standard washing up liquid in line with local cleaning and hygiene policies in
place.
Disconnect the pump from the mains before cleaning. To avoid electrical shock, never clean the pump with AC power adapter plugged
into an outlet or when pump is on.
Do not immerse the pump in water.
Do not use aggressive cleaners or disinfectants such as acetone, benzene or iodine.
The pump is not intended to be cleanevd using a dishwasher.
Do not sterilize the pump, e.g. by irradiation, EtO, steam or in an autoclave.
Before using or setting a feeding regimen, check the integrity of the pump, the AC power plug and its accessories. Do not use the pump if it is damaged.
ACCESSORIES
Dedicated carry bags and pump stands are available for the Flocare Infinity III pump. Please contact your local Nutricia sales representative for availability and
ordering information of the respective parts.
RECYCLING & DISPOSAL
The pump and its electrical accessories (AC power plug, PDMS cable, Nurse Call cable), should be disposed according to local standards and regulations governing
the disposal of electronic waste (e-waste). Other accessories can be disposed of or recycled as standard non-hazardous waste.
Note: In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number.
11.
SPECIFICATIONS [e-IFU]
11.1.
PERFORMANCE
Volumetric accuracy
Battery life
Air bubble detection
Upstream occlusion pressure
Downstream occlusion pressure
11.2.

OPERATING

Atmospheric pressure
Temperature
Humidity
11.3.
STORAGE
Atmospheric pressure
Temperature
Humidity
Shelf life
Note: Time to recover from maximum and minimum storage temperature to operating temperature is ~30 min.
ZMNL900901_DE.indb 10
ZMNL900901_DE.indb 10
] can provide suggestions on the display to solve an alarm or warning condition.
±5%
~24h @ 125mL/h, ~12h @ 250mL/h
> 1.5 mL of air in giving set after the pump set
-34 kPa ± 21 kPa for 5 sec.
84 kPa ± 21 kPa for 30 sec. / 103 kPa ± 21kPa for 5 sec.
65 – 106 kPa
5° – 40°C
15 – 95%, non-condensing
50 – 106 kPa
-20° – 60°C
10 – 95%, non-condensing
None. A normal expected use life is 5 years. The battery life degrades over time and should be charged before any use
] button is the only button that acknowledge and dismiss an alarm or warning
‒ 10 ‒