7.4.
SETTINGS MENU [e-IFU]
tings.
'Define Therapy list' controls the available therapies that are visible when the THERAPY [
with a green switch symbol.
7.4.1. Reset patient data
Reset patient data is designed to erase all collected patient data (non-personal)
The following data will reset: Therapy parameters, Progress bar, Patient history data, Personalised therapies, Define therapy list, Sound settings and Display settings.
7.4.2. Sound [e-IFU]
The Flocare Infinity III supports 3 AUDIO VOLUMES, 'Low, Medium and High' and can be set by the user.
For safety reasons the alarm volume will be automatically escalated to High in both settings 'Low' and 'Medium' after 3 min. This safety feature cannot be turned off.
The THERAPY COMPLETED ALARM is activated by default and the pump will generate an audio and visual warning. If it is not desired to be disturbed by the audio
warning, the THERAPY COMPLETED ALARM can be switched OFF.
7.5.
PROGRAMMING THE PUMP
Caution: Before programming the pump, always be convinced that your selected therapy and its parameters are correct. In case of any doubts, consult your care giver.
Therapies are programmed via the THERAPY [
Volume – Rate
Rate
the
Intermittent
IMPORTANT:
Therapies RATE and INTERMITTENT are disabled by default, if a desired therapy does not appear when the THERAPY button is pressed, you can enable
be
them by selecting: SETTINGS DEFINE THERAPY LIST. [e-IFU]
7.5.1. Program a VOLUME - RATE therapy
test
•
From the home screen press the THERAPY [
•
Select VOLUME – RATE by pressing the associated soft button.
and
•
Press the VOLUME soft button
•
Change the Volume using the '+' and '-' buttons and press the '>' button to accept the value.
•
Press the RATE soft button
•
Change the Rate using the '+' and '-' buttons and press the '>' button to accept the value.
S
•
Check your total therapy time underneath the VOLUME or RATE panel for additional safety.
•
Press the START/PAUSE [
7.5.2. Program a RATE therapy (not recommended for volume sensitive patients)
•
From the home screen press the THERAPY [
•
Enter the RATE soft button
•
Change the Rate using the '+' and '-' buttons and press the '>' button to accept the value.
•
Press the START/PAUSE [
7.6.
LAST FEEDING INCOMPLETE
When a therapy has not been administered completely whilst the pump has been turned OFF, the pump will show the warning message: 'Last feeding incomplete'.
You either have the option to RESTART the active therapy where the progress bar will be reset or you can RESUME the delivery of the active therapy.
7.7.
INFORMATION & MEMORY FUNCTIONS [e-IFU]
7.7.1. 30-day history graph
The Flocare Infinity III stores all administered therapies for a maximum period of 30 days.
The daily administered volume can be accessed by pressing SOFT BUTTON 1 from the home screen and pressing HISTORY. With the arrow buttons, you can
navigate to earlier or later days. By pressing the + magnifying symbol you can zoom in on your selected day by showing a table of individual therapies.
are
7.7.2. Reset progress bar
The Progress bar is automatically cleared after completion of each therapy where a VOLUME was programmed.
When there is a need to manually clear the Progress bar, navigate from the HOME screen to SOFT BUTTON 1 RESET PROGRESS BAR.
7.8.
BATTERY [E-IFU]
vate
The Flocare Infinity III is equipped with a 3.7V, 3200 mAh Lithium Ion battery and needs to be fully charged for a period of 6 hours prior to or during first use, this will
maximise the battery life time.
ped
Note: The battery capacity is an estimate. If you are unsure that enough battery capacity remains for your intended therapy, connect to the AC power adapter until
recharged.
ZMNL900901_DE.indb 9
ZMNL900901_DE.indb 9
] button. The Flocare Infinity III supports the following therapies:
Administering of a programmed VOLUME (mL) with a programmed RATE (mL/h).
Administering with a programmed RATE (mL/h) but without a programmed VOLUME (mL)
Administering of a programmed VOLUME (mL) divided over different portions with a programmed RATE (mL/h) with programmed PAUSE
(interval) times. [e-IFU]
] button
] button to start your VOLUME-RATE therapy.
] button
] button to start your RATE therapy.
] button is pressed. Therapies that are currently enabled are marked