2

Intended use

The BiCision instrument is intended for bipolar coagulation with
HF current and for mechanical division of tissue during open and
laparoscopic procedures.
3

Normal use

This product is used with the following Erbe electrosurgical
units:
• VIO 300 D (with BiClamp upgrade)
• VIO 3
4

Maximum electrical capacity

The maximum electrical capacity of this product is:
• 250 V
p
5

Safety instructions

WARNING!
This product may be used only by trained medical staff, who
have been shown how to use it according to the instructions for
use.
Do not use the product on vessels with a diameter greater than
7 mm.
Do not use the product for tubal sterilization or for tubal coagu-
lation for sterilization purposes, because there is no evidence of
the suitability of the product as yet.
Check the product for damage before use, including all the insu-
lation.
If damaged, do not use this product!