2. Repeat the coagulation and cutting procedure as previously
described on another tissue site as required.

6.14 Removing the instrument from a trocar

1. For laparoscopic procedures: Remove the instrument from
the trocar with the instrument jaws closed.

6.15 Removing contamination from the instrument jaws

To ensure the proper function of the instrument, the instrument
jaws must be cleaned at regular intervals (e.g. in breaks during
use) to remove any significant contamination.
1. If necessary, remove the instrument (with the instrument
jaws closed) from the trocar.
2. Use the included cleaning implement to remove any coarse
contamination on the instrument jaws:
– Use the cup-shaped scraper of the cleaning implement
to clean the hollow of the bottom jaw.
– Use the sharp spatula of the cleaning implement to clean
the crevice of the upper jaw.