1a
3. Hold the electrode by the insulation (1a) between thumb and
forefinger.
Screw the electrode with the rotary wheel (3a) tightly to the
stop (about 6 revolutions) without cross-threading it.
4. Hold the handle (4) like a pen so the grippers point upwards
and the movable grippers are not blocked and press the but-
ton.
Slide the shaft tube straight down until it clicks into the han-
dle (4).
The product is correctly mounted when the rotary wheel can be
rotated 360° and you can open and close the instrument jaws
without difficulty using the handle.
5. Attach the product to the matching Erbe connecting cable
(20196-045, 20196-053, 20196-055, 20196-057, 20196-
061, 20196-062, 20196-063 or 20196-064).
6. Connect the product to the bipolar socket.